EO or gamma sterilisation: what should importers of medical disposables ask for?

Buyer guide, updated 2026-11-25

In brief

  • Ethylene oxide (EO) is a gas process validated to ISO 11135, and radiation, including gamma, is validated to the ISO 11137 series.
  • Both should reach a sterility assurance level (SAL) of 10^-6 for a device labelled sterile.
  • EO leaves residues, so ask for residual EO and ethylene chlorohydrin results against ISO 10993-7.
  • Radiation can damage some polymers, such as PTFE and natural polypropylene, and can yellow PVC, so ask for material compatibility data.
  • For every lot, ask for a sterilisation certificate that names the method, the cycle or dose, and the evidence used to release the lot.

Ethylene oxide (EO) and gamma radiation are both accepted ways to sterilise single-use medical devices, and neither is better in general. The choice depends on the materials and packaging of the product. An importer should ask which method was used, which standard the process is validated to (ISO 11135 for EO, ISO 11137 for radiation), and for lot records that show the validated cycle or dose was delivered.

How each method works

Ethylene oxide is a gas. The US Centers for Disease Control and Prevention (CDC) describes a basic cycle with five stages: preconditioning and humidification, gas introduction, exposure, evacuation and air washes. It gives typical operating ranges of 450 to 1200 mg/l gas, 37 to 63°C, 40 to 80% relative humidity and 1 to 6 hours of exposure. EO kills microorganisms by alkylation of protein, DNA and RNA. Because many materials absorb EO, products must be aerated afterwards so the gas can desorb. The CDC gives 8 to 12 hours of mechanical aeration at 50 to 60°C in a healthcare setting.

Gamma radiation penetrates the sealed carton. No gas has to reach the device, so packaging need not be gas-permeable, and there is no chemical residue to aerate. The dose absorbed by the product, measured in kilograys (kGy), is what matters. ISO 11137-2 gives methods to set the minimum dose needed and to substantiate 25 kGy or 15 kGy as the sterilisation dose for an SAL of 10^-6.

EO and gamma compared

Point Ethylene oxide (EO) Gamma radiation
Process standard ISO 11135:2014, with Amd 1:2018 (revised Annex E, single batch release) ISO 11137-1:2025 (requirements), ISO 11137-2:2013 with Amd 1:2022 (dose), ISO 11137-3 (dosimetry)
What kills microbes Alkylation of protein, DNA and RNA by the gas Absorbed radiation dose
Packaging Must be sealed in a gas-permeable package (FDA) Gas permeability not needed
Residues EO and ethylene chlorohydrin (ECH); limits in ISO 10993-7 No chemical residue
Aeration Needed Not needed
Material cautions Suits heat- and moisture-sensitive items PTFE significantly damaged; natural polypropylene loses properties; PVC yellows above about 30 kGy
Routine release evidence Recorded cycle parameters and, where used, biological indicator results Dosimeter readings showing the minimum dose was met and the maximum not exceeded
Lot label symbol (ISO 15223-1) 5.2.3 "Sterilized using ethylene oxide" (STERILE EO) 5.2.4 "Sterilized using irradiation" (STERILE R)

Which materials suit which method

The US FDA says devices made from certain polymers, metals or glass, devices with several packaging layers, and devices with hard-to-reach areas, such as catheters, are likely to be EO-sterilised. It also says EO is sometimes the only method that sterilises a device without damaging it. The FDA cites literature indicating that about half of all sterile medical devices in the US are sterilised with EO.

For radiation, the irradiation service provider Sterigenics publishes polymer ratings. It states that PTFE and PFA are significantly damaged when irradiated. Natural polypropylene has its physical properties greatly reduced, while stabilised grades rate better. PVC shows significant yellowing at doses above 30 kGy, polystyrene begins to yellow above 50 kGy, and medium-density and linear low-density polyethylene grades are preferred over high-density.

For an importer, the question is: was the method chosen after material testing at the maximum dose or the worst-case cycle, and can the supplier show that data?

Sterility assurance level

The FDA's 510(k) sterility guidance (January 2024) asks for an SAL of 10^-6 for devices labelled sterile, and 10^-3 only for devices that contact intact skin. In Europe, EN 556-1 requires a probability of a viable microorganism of 10^-6 or less before a terminally sterilised device may be designated "STERILE". If a supplier quotes an SAL other than 10^-6 for an invasive product, ask why.

Residues after EO

The FDA says limits on residual EO and ECH matter because long-term exposure to EO has been linked to cancer. ISO 10993-7 sets allowable limits for both. A third edition, ISO 10993-7:2026, was published in April 2026; the FDA's sterilisation page still lists the 2008 edition (reaffirmed 2012) as the recognised version. Ask the supplier which edition its residue testing follows, which contact-duration category it applied to the product, and for the measured EO and ECH results against the limits.

How lots are released

  • Release with biological indicators (BI). A BI is an inoculated carrier with a defined resistance to the process, made to the ISO 11138 series. BIs placed in a process challenge device are incubated after the EO cycle, and the lot is released when none grow.
  • Parametric release. Product is declared sterile from physical and chemical process data, such as time, temperature, humidity and gas concentration, rather than from sample testing. It is allowed only when validation supports it.
  • Dosimetric release (radiation). Dosimeters at reference positions fixed by dose mapping show the dose delivered to each load. The lot is released when the calculated minimum dose meets the sterilisation dose and the maximum stays within the limit tested for the materials. ISO 11137-2 also requires periodic dose audits to confirm the dose stays effective.

Shelf life and re-sterilisation

The use-by date (ISO 15223-1 symbol 5.1.4) depends on the sterile barrier staying intact, so ask for packaging and ageing data behind the stated shelf life. Do not re-sterilise single-use devices. The "Do not resterilize" symbol (5.2.6) is there because a second EO cycle adds residues and a second radiation dose adds to material damage. Sterigenics rates re-sterilisation as "not likely" for PVC, natural polypropylene and PTFE.

How to check a supplier's sterilisation

  1. Ask who sterilises the product; if a contract site, ask for its name and certificate.
  2. Ask for the validation summary: standard and edition, validation method, cycle parameters or sterilisation dose, and the date of the last revalidation or dose audit.
  3. Check the declaration of conformity and the technical file name the same method and standard as the label.
  4. Ask for EO and ECH results (EO products) or maximum-dose material data (radiation products).
  5. Match the lot number on the sterilisation certificate to the CoA, the label and the packing list.

Worked example

An importer orders four lines from Vibama Surgicals in one container:

  • Disposable syringe, luer slip, REF INJ-001, 1 to 50 ml, ISO 7886-1, labelled STERILE EO, HS 9018.31.
  • IV infusion (giving) set, REF INF-003, 20 drops/ml, ISO 8536-4, labelled STERILE EO, HS 9018.90.
  • Petri dish, REF LAB-001, 90 mm, single or triple vent, labelled STERILE R, HS 3926.90. It is laboratory ware, not a medical device in India.
  • Biological indicator vials, REF STE-007, for steam or EO, ISO 11138, for the importer's own hospital customers.

The importer's document request lists, for INJ-001 and INF-003: a sterilisation certificate per lot naming ISO 11135, the cycle reference and release basis (BI results or parametric release); EO and ECH residue results against ISO 10993-7; and a CoA per lot. For LAB-001: a sterilisation certificate per lot showing the minimum and maximum dose delivered, with the sterilisation dose and date of the last dose audit. For STE-007: the CoA with organism, population and resistance data per lot. For all lines: the packing list with lot and expiry per carton. See quality and documents for what is available, and the syringes and needles and laboratory plastics ranges.

Frequently asked questions

Is gamma sterilisation better than EO?

No. Each suits different products. Gamma leaves no chemical residue but can damage some polymers, such as PTFE, while EO suits heat- and moisture-sensitive items but needs aeration and residue testing.

What does STERILE EO on a label mean?

It is the ISO 15223-1 symbol 5.2.3 and indicates the device was sterilised using ethylene oxide. STERILE R (symbol 5.2.4) indicates sterilisation by irradiation.

Which standard sets EO residue limits?

ISO 10993-7 sets allowable limits for residual ethylene oxide and ethylene chlorohydrin. The current edition was published in April 2026, so check which edition the supplier's test report cites.

Is 25 kGy always the gamma dose?

No. ISO 11137-2 gives methods to set a product-specific minimum dose and to substantiate 25 kGy or 15 kGy for an SAL of 10^-6. Ask for the dose the validation established.

Who manufactures sterile medical disposables in India for export?

Vibama Surgicals, in Surat, India, makes 427 products in 28 departments under a medical device manufacturing licence issued under India's Medical Devices Rules, 2017, with an ISO 13485 certified quality system and CE-marked products. Sterile lines include the disposable syringe (REF INJ-001, STERILE EO), the IV infusion set (REF INF-003, STERILE EO) and the Petri dish (REF LAB-001, STERILE R), with a CoA and sterilisation certificate per lot. See quality and documents or contact the export team.

Sources