Surgicals

How should importers specify IV cannulas? Gauges, colour codes, ISO 10555-5 and export documents

Product guide, updated 2026-11-17

In brief

  • Peripheral IV cannulas (over-needle catheters) are made and tested to ISO 10555-5, with the general rules in ISO 10555-1.
  • The hub colour shows the size: 14G orange, 16G grey, 17G white, 18G green, 20G pink, 22G blue, 24G yellow.
  • A tender-ready specification states gauge, length, catheter material, port or no port, wings, safety mechanism and flow rate.
  • ISO 10555-5 requires the manufacturer to state flow rate, colour code and outside diameter, and to warn against re-inserting a withdrawn needle.
  • IV cannulas normally ship under HS heading 9018.39 (catheters, cannulae and the like).

An IV cannula is specified by its gauge (shown by the hub colour), length, catheter material, whether it has an injection port and wings, and whether it has a needle-stick safety mechanism. Peripheral IV cannulas are made and tested to ISO 10555-5, which sets the colour code, the performance tests and what the manufacturer must print on the pack. A buyer who sends those details gets comparable quotes and avoids rejected lots at tender evaluation.

The standard: ISO 10555-5

ISO 10555-5:2013 covers over-needle peripheral intravascular catheters supplied sterile for single use, and is read together with ISO 10555-1 for general requirements. Among other things it requires:

  • hubs or fittings colour coded to show the nominal outside diameter;
  • tests including needle-hub strength and vent-fitting leakage;
  • the manufacturer to state the flow rate for each lumen, warn against re-inserting a partly or fully withdrawn needle, and print the colour code and outside diameter on each primary pack unless visible.

A revision is in progress (ISO/DIS 10555-5:2026), so check which edition a tender cites.

Gauge, colour and typical flow

Gauge Hub colour Approx. outside diameter Indicative gravity flow Typical use
14G Orange 2.1 mm 250–330 ml/min Trauma, rapid volume
16G Grey 1.7 mm 180–220 ml/min Major surgery, rapid transfusion
17G White 1.5 mm 125–150 ml/min Surgery where 16G is too large
18G Green 1.3 mm 90–120 ml/min Blood transfusion, contrast
20G Pink 1.1 mm 60–65 ml/min Routine adult infusion
22G Blue 0.9 mm 30–36 ml/min Children, fragile veins
24G Yellow 0.7 mm 13–20 ml/min Neonates, small veins

Flow figures vary by catheter length and design; the binding figure is the one the manufacturer states for its own product under ISO 10555-5. Most public hospital tenders order 18G, 20G, 22G and 24G in the largest volumes.

What to put in the specification

Field Options to state
Gauge and length e.g. 20G × 32 mm
Catheter material FEP or PTFE (radiopaque stripes), or polyurethane
Injection port With port (valve and cap) or without
Wings Winged or wingless
Safety mechanism Standard, or passive safety (needle shield) for programmes that require sharps protection
Connector Luer lock to ISO 80369-7
Sterilisation Ethylene oxide, with residue limits
Packing Units per dispenser box and per carton; label language

Testing and documents per lot

Ask for a certificate of analysis per lot covering at least dimensions, flashback, needle-hub strength, leakage, flow rate and sterility, plus the sterilisation certificate and ethylene oxide residue results. For registration, the regulator will usually also want the declaration of conformity, the product specification and the labels as printed.

Checking a new supplier

Before a first order, compare suppliers on evidence rather than on price per piece:

  • Samples from a production lot, not hand-picked units, with that lot's certificate of analysis. Check that the hub colours match the ISO code, the catheter tip is smooth with no flaring, and the needle withdraws without dragging.
  • Flashback and flow. Insert a sample into a water-filled tube and watch for immediate flashback; time the flow against the figure printed on the pack.
  • Sterilisation evidence. For ethylene oxide, ask for the cycle certificate for the lot and the residue result. Shelf life claims, usually printed as a use-by date, should be backed by real-time or accelerated ageing data.
  • Pack integrity. Peel the blister: the seal should open cleanly along the chevron without tearing the paper, and the lot and expiry should be legible on every unit.
  • Registration status. Confirm the manufacturer can supply the documents your regulator lists, in particular a free sale certificate and a declaration of conformity naming ISO 10555-1 and ISO 10555-5.

A supplier that sends lot-matched samples, test results and a clear label proof in your language within a few days is usually one that can also hold delivery dates on a tender.

HS code and customs

IV cannulas are normally classified under HS heading 9018.39, catheters, cannulae and the like. Confirm the full national tariff line with your customs broker, since some countries split 9018.39 further.

Worked example: a tender for four sizes

A ministry tender asks for 1,000,000 cannulas: 300,000 each of 18G and 20G, and 200,000 each of 22G and 24G, with injection port and wings, packed in 50s.

  • Product: IV cannula with injection port, ref INF-001, ISO 10555-5.
  • Packing: 50 per dispenser box gives 20,000 boxes in total. The carton count follows the tender's own units-per-carton figure.
  • Usually ordered alongside: IV infusion giving sets (ISO 8536-4) and safety IV cannulas where the tender asks for sharps protection.
  • Documents with the bid: free sale certificate, declaration of conformity, a recent certificate of analysis for each gauge, and the sterilisation certificate.

Frequently asked questions

Which standard applies to IV cannulas?

ISO 10555-5 for over-needle peripheral intravascular catheters, read with ISO 10555-1 for general requirements. Many national standards are identical adoptions, such as EN ISO 10555-5:2013 and BS EN ISO 10555-5:2013.

What colour is a 20G IV cannula?

Pink. 18G is green, 22G blue and 24G yellow, under the ISO colour code for catheter hubs.

What must be printed on an IV cannula pack?

ISO 10555-5 requires the colour code and outside diameter on each primary pack unless visible, a stated flow rate, and a warning not to re-insert a withdrawn needle, in addition to the lot, expiry and sterility information that regulators require.

What is the HS code for IV cannulas?

HS 9018.39, catheters, cannulae and the like. The national tariff line under it varies by country.

Who manufactures IV cannulas in India for export?

Vibama Surgicals supplies IV cannulas from 14G to 26G with or without injection port, plus safety IV cannulas from 18G to 24G, made to ISO 10555-5 under an ISO 13485 certified quality system and sterilised by ethylene oxide, with private-label packing. See the infusion and IV range or request a quote.

Sources

Request a quote

List the items, sizes and quantities, and the port you want delivery to. We reply within two working days with prices, lead time and the documents we will send.

Request a quoteAsk on WhatsApp