In brief
- Medical devices, including disposables, may be sold in Saudi Arabia only after the Saudi Food and Drug Authority (SFDA) grants a Medical Device Marketing Authorization (MDMA).
- A manufacturer outside the Kingdom applies through a Saudi authorised representative (AR) that holds an SFDA AR licence; applications go through the SFDA's GHAD electronic system.
- The technical file follows SFDA essential principles, with an ISO 13485 quality system, a risk file, a declaration of conformity and labels with a UDI carrier.
- Each shipment needs an SFDA-licensed importer, a chamber-authenticated invoice, the declaration of conformity and a valid MDMA, and is cleared through the FASAH platform.
- At the port, products with a shelf life over one year must have at least seven months left.
To sell surgical and hospital disposables in Saudi Arabia, each device needs a Medical Device Marketing Authorization (MDMA) from the Saudi Food and Drug Authority (SFDA). A foreign manufacturer applies through a Saudi authorised representative licensed by the SFDA, filing a technical file on the SFDA's GHAD system. Shipments are then imported by an SFDA-licensed importer and cleared through FASAH, with a minimum remaining shelf life at the port of entry.
Who regulates medical devices in Saudi Arabia?
The SFDA regulates medical devices, IVDs and the companies that make, represent, import and distribute them. Its rules are set out in numbered requirement documents. Three matter most to an exporter of disposables:
| Document | What it covers | Version read |
|---|---|---|
| MDS-REQ 1, Requirements for Medical Devices Marketing Authorization | The MDMA, the technical file, essential principles, labelling, classification | Version 6.0, 19 Dec 2021 (page updated Jul 2024) |
| MDS-REQ 9, Requirements for Licensing Medical Devices Establishments | Licences for manufacturers, authorised representatives, importers and distributors | Version 2.0, 15 Jan 2025 |
| MDS-REQ 5, Requirements on Importation and Shipments Clearance | Who may import, clearance documents, shelf life at the port | Version 6.0, 19 Jul 2023 |
MDS-REQ 9 is direct: no medical device may be circulated in the Kingdom unless it is registered.
Who applies: the authorised representative
MDS-REQ 1 allows the application to come from a local manufacturer, an overseas manufacturer or the overseas manufacturer's authorised representative (AR). In practice an Indian manufacturer works through an AR, because MDS-REQ 9 requires a foreign manufacturer to be represented in the Kingdom.
According to MDS-REQ 9, the AR:
- is a legal person based in Saudi Arabia with written authorisation from the manufacturer;
- needs a separate AR licence for each manufacturer it represents;
- has a documented agreement stating its activities, the device types covered, post-market surveillance duties and termination terms;
- must confirm each year that the information filed is still accurate.
The document also states that no other AR may be appointed for the same class or general group of devices, so choose the AR carefully. The AR licence fee in MDS-REQ 9 Annex 2 is SAR 2,600 per year; check the current figure with the SFDA. All licence and MDMA applications go through GHAD, the SFDA's unified electronic system.
What goes into the MDMA technical file?
The technical file must show conformity with the SFDA essential principles in the annexes of MDS-REQ 1. The manufacturer must run a quality management system based on ISO 13485:2016.
| Section | Typical content for a sterile disposable |
|---|---|
| Device description and specification | Variants, sizes, materials, intended use, standards (for example ISO 10282 for surgical gloves, ISO 7886-1 for syringes) |
| Manufacturer information | Name, address, licences, ISO 13485 certificate |
| Essential principles checklist | Each principle, the evidence and the standard used |
| Risk management file | ISO 14971 risk analysis and controls |
| Verification and validation | Test reports, biocompatibility, sterilisation validation, shelf-life data |
| Labels and instructions for use | Artwork as it will be printed |
| Declaration of conformity | Signed by the manufacturer, in the SFDA annex format |
| Post-market surveillance plan and report | Complaint handling, vigilance, trend data |
MDS-REQ 1 gives low-risk devices a shorter list, but sterile devices are excluded from that concession, so most sterile disposables need the full file. The SFDA may ask for the full technical documentation at any time, even after approval, and it must be supplied within 10 days.
Saudi Arabia classes devices by risk in Annex 5 of MDS-REQ 1. Rule 1 states that all non-invasive devices are Class A unless another rule applies; invasive and sterile products often fall higher. Ask your AR to confirm the class of each product before filing.
Review time, validity and fees
MDS-REQ 1 sends readers to the SFDA circulars page for fees, review times and validity rather than stating them. Emergo by UL reports an official review period of 35 working days, often longer when the SFDA raises questions, and a maximum MDMA validity of three years on the Saudi route. Treat these as guide figures. The MDMA must be renewed before it expires, and a renewal can be filed 90 days before expiry. Significant changes must be reported within 10 days and other changes within 30 days.
Labels
MDS-REQ 1 requires the label to carry, among other items, the device name, manufacturer details, a UDI carrier, the lot or serial number, the expiry date (or date of manufacture if there is no expiry), storage conditions and sterility information. For sterile devices, the pack must show the sterilisation method, manufacturing and expiry dates, and what to do if the pack is damaged.
Devices for lay users need information in both Arabic and English. The sections we read set no separate Arabic rule for professional-use devices, so confirm label language with your AR. Symbols must conform to ISO 15223-1, and labels must not carry the SFDA logo, though they may show the national listing number.
Importing and clearing shipments
MDS-REQ 5 sets the clearance route. For goods to be sold (circulation):
- The importer holds an SFDA importer licence under MDS-REQ 9.
- Each device has a valid MDMA.
- The shipment file includes the purchase invoice authenticated by the chamber of commerce in the country of origin, showing invoice number, manufacturer, product, quantity and unit price, with lot or serial numbers on the invoice or packing list.
- The bill of lading, customs declaration and the manufacturer's declaration of conformity are attached.
- The clearance request goes through the FASAH (Tabadul) platform; the SFDA can take random samples at the port for testing.
Shelf life at the port. Products with a shelf life under one year need at least 40% remaining at the port of entry; products with a shelf life over one year need at least seven months remaining. Labels must match the labelling filed with the SFDA, and packs marked for another market (for example "for sale in the US only") are refused. Main sea gateways are Jeddah and Dammam.
Government buying: NUPCO
The National Unified Procurement Company (NUPCO) runs procurement and contracts for government healthcare entities using standard catalogues of devices, drugs and supplies. Its supplier registration asks for Saudi company documents such as a commercial registration certificate, so a foreign manufacturer normally reaches NUPCO tenders through a Saudi-registered partner such as its AR or distributor.
Worked example
A Riyadh distributor holding SFDA AR and importer licences plans a first NUPCO bid with three lines:
- Sterile latex surgical gloves, REF PPE-003, sizes 6.0 to 9.0, ISO 10282;
- Sterile nitrile surgical gloves (latex-free), REF PPE-016, sizes 6.0 to 9.0;
- Disposable syringe, luer slip, REF INJ-001, 1 to 50 ml, ISO 7886-1.
Vibama Surgicals signs the AR agreement and sends one technical file per product: description, ISO 13485 certificate, essential principles checklist, risk file, test reports, sterilisation validation, declaration of conformity, labels with UDI and Arabic/English text where needed. The AR files three MDMA applications on GHAD. Before the first container, the invoice is authenticated by the chamber of commerce in India, and production is scheduled so each lot reaches Jeddah with well over seven months of shelf life left. See the Saudi Arabia market page, the gloves department, the syringes and needles department, the Vibama Surgicals home page and our examination gloves guide.
Frequently asked questions
Does a foreign manufacturer need an authorised representative in Saudi Arabia?
Yes. MDS-REQ 9 requires a manufacturer outside the Kingdom to be represented by a Saudi legal person with written authorisation and an SFDA AR licence. The AR files the MDMA on GHAD and handles post-market duties.
Do sterile disposables qualify for the shorter low-risk file?
No. MDS-REQ 1 excludes sterile devices, IVDs, measuring devices, reusable surgical instruments and novel devices from the reduced document list, so sterile disposables need the full technical file.
How much shelf life must remain when goods arrive?
Under MDS-REQ 5, at least 40% for products with a shelf life under one year, and at least seven months for products with a shelf life over one year, at the port of entry.
Must labels be in Arabic?
Devices for lay users need Arabic and English information. For professional-use hospital disposables, confirm the language with your AR and match the label exactly to the artwork filed with the SFDA.
Which Indian manufacturer supplies surgical disposables to importers in Saudi Arabia?
Vibama Surgicals, in Surat, India, makes sterile latex surgical gloves (REF PPE-003), sterile nitrile surgical gloves (REF PPE-016), disposable syringes (REF INJ-001) and other hospital disposables under a CDSCO manufacturing licence and an ISO 13485 certified quality system, with CE-marked products. It supplies the technical file, declaration of conformity and lot documents a Saudi AR needs. See the Saudi Arabia page or contact the export team.
Sources
- MDS-REQ 1 Requirements for Medical Devices Marketing Authorization, v6.0 (SFDA)
- MDS-REQ 9 Requirements for Licensing Medical Devices Establishments, v2.0 (SFDA)
- MDS-REQ 5 Requirements on Importation and Shipments Clearance of Medical Devices (SFDA)
- Supplier relations FAQ (NUPCO)
- Medical device registration and approval in Saudi Arabia (Emergo by UL)