Surgicals

How do you specify sterile surgical gloves to ISO 10282?

Product guide, updated 2026-11-26

In brief

  • Sterile single-use surgical gloves are specified by ISO 10282:2023, the fourth edition, which replaced ISO 10282:2014.
  • Sizes run from 5 to 9.5 in half sizes, each with a set palm width and minimum length; minimum single-wall thickness is 0.10 mm on smooth areas.
  • Watertightness (freedom from holes) is sampled at inspection level G-I with AQL 1.5; dimensions and tensile properties at S-2, AQL 4.0.
  • Powder-free gloves may carry at most 2.0 mg of powder per glove; powdered gloves at most 15 mg/dm².
  • Natural rubber latex gloves are Type 1 and synthetic gloves (such as nitrile or polyisoprene) are Type 2, with different tensile limits.

Sterile surgical gloves are specified by ISO 10282:2023, "Single-use sterile rubber surgical gloves: Specification". It sets sizes, minimum thickness, watertightness at AQL 1.5, force and elongation at break for latex and synthetic gloves, powder limits, sterility and package marking. A good tender names this standard, the material, the finish, the sizes and the documents needed with each lot.

What does ISO 10282 cover?

ISO describes ISO 10282 as specifying packaged sterile rubber gloves for use in surgical procedures to protect the patient and the user from cross-contamination. It covers single-use gloves worn once and discarded, smooth or textured. It does not cover examination or procedure gloves, which follow ISO 11193 and EN 455, and it does not set sterilisation, handling or storage procedures.

The fourth edition was published in February 2023 and replaced ISO 10282:2014. The foreword lists the main changes: new figures for measuring dimensions (Annex A), updated definitions of powdered and powder-free surfaces, a statement on testing gloves more than six months old, and a new Annex A, with the former Annex A renamed Annex B. India adopted it unchanged as IS 13422:2024.

The standard divides gloves by:

  • Material: Type 1, natural rubber latex; Type 2, synthetic rubber latex or solution (for example nitrile or polyisoprene).
  • Design: straight fingers or curved fingers.
  • Finish: smooth or textured; powdered or powder-free.

Sizes and dimensions

Size Palm width (mm) Minimum length (mm)
5 67 ± 4 250
5.5 72 ± 4 250
6 77 ± 5 260
6.5 83 ± 5 260
7 89 ± 5 270
7.5 95 ± 5 270
8 102 ± 6 270
8.5 108 ± 6 280
9 114 ± 6 280
9.5 121 ± 6 280

Minimum single-wall thickness is 0.10 mm on smooth areas and 0.13 mm on textured areas, for every size. Gynaecological or long-cuff gloves are bought to the same standard with a longer cuff stated in the specification.

Sampling and AQL

Property Inspection level AQL
Dimensions (width, length, thickness) S-2 4.0
Watertightness (freedom from holes) G-I 1.5
Force and elongation at break, before and after ageing; force at 300% elongation S-2 4.0

Watertightness is tested by filling the glove with 1,000 cm³ ± 50 cm³ of water and checking for leaks immediately and again after two to four minutes. Some buyers ask for a tighter AQL on holes than 1.5; if so, write it into the tender as a buyer requirement, not as "ISO 10282".

Tensile properties

Property Type 1 (latex) Type 2 (synthetic)
Force at break before ageing, minimum 12.5 N 9.0 N
Elongation at break before ageing, minimum 700% 600%
Force at 300% elongation before ageing, maximum 2.0 N 3.0 N
Force at break after ageing, minimum 9.5 N 9.0 N
Elongation at break after ageing, minimum 550% 500%

Accelerated ageing is 168 hours at 70 °C. Gloves older than six months, or of unknown manufacturing date, need only meet the after-ageing values.

Powdered or powder-free?

ISO 10282 limits powder-free gloves to 2.0 mg of powder residue per glove and powdered gloves to 15 mg/dm². Surface-powdered gloves must carry a warning to remove the powder aseptically before surgery. Powdered surgeon's gloves are banned in the United States under an FDA rule published on 19 December 2016, and many hospital tenders elsewhere now ask for powder-free only, so check before quoting powdered gloves.

Latex-sensitive patients and staff need a Type 2 glove. Type 1 gloves must carry a natural rubber latex allergy warning on the unit package.

Marking

The unit package must show the manufacturer or trademark, material, design (straight or curved fingers), finish (textured or smooth; powdered or powder-free), size, lot number, date of manufacture (year and month), "STERILE UNLESS THIS PACKAGE IS OPENED OR DAMAGED", "FOR SINGLE USE", "SURGICAL GLOVES", the expiry date and, for Type 1, the latex allergy warning. The inner wrap shows the size and left and right hands. Gloves must be sterilised, and the method must be disclosed on request.

What to write in a tender specification

  1. Product: sterile single-use surgical gloves, pair packed, to ISO 10282:2023.
  2. Material: Type 1 natural rubber latex, or Type 2 synthetic (state nitrile, polyisoprene or other).
  3. Finish: powder-free (2.0 mg per glove maximum), textured or smooth; design: curved or straight fingers.
  4. Sizes and split, for example 6.5, 7, 7.5 and 8.
  5. Watertightness AQL 1.5 or tighter, and minimum thickness.
  6. Sterilisation: gamma or EO, with a sterilisation certificate per lot.
  7. Shelf life and minimum remaining shelf life on delivery.
  8. Packing: inner wrap, peel pouch, dispenser box, export carton; label language.
  9. Documents: certificate of analysis per lot, sterilisation certificate, declaration of conformity, ISO 13485 certificate, free sale certificate.

HS code and documents per lot

Surgical gloves of vulcanised rubber classify under HS 4015.12 (gloves, mittens and mitts of a kind used for medical, surgical, dental or veterinary purposes). Confirm the full national tariff line with your broker. Each lot should ship with a certificate of analysis showing dimensions, watertightness, tensile results and powder residue, plus the sterilisation certificate.

How to check a supplier

Ask for the ISO 10282 test report from an independent laboratory, the ISO 13485 certificate and its scope, the manufacturing licence listing surgical gloves, sterilisation validation, and samples from a production lot. Test a few boxes yourself for holes and fit before the first order.

Worked example

A hospital group tenders for theatre gloves in three lines:

The specification names ISO 10282:2023, AQL 1.5 for watertightness, powder-free at 2.0 mg per glove maximum, gamma sterilisation and a certificate of analysis per lot. For examination gloves on the same tender, see our examination gloves guide, and the gloves and protective wear department for the full range, or start from the Vibama Surgicals home page.

Frequently asked questions

What is the current edition of ISO 10282?

ISO 10282:2023, the fourth edition, published in February 2023. It replaced ISO 10282:2014.

What AQL does ISO 10282 set for holes?

Watertightness is sampled at inspection level G-I with an AQL of 1.5. Dimensions and tensile properties use inspection level S-2 with AQL 4.0.

What sizes do surgical gloves come in?

ISO 10282 lists sizes 5 to 9.5 in half sizes, each with a palm width and minimum length. Size 7 has a width of 89 ± 5 mm and a minimum length of 270 mm.

How much powder can a powder-free surgical glove contain?

No more than 2.0 mg of powder residue per glove under ISO 10282. Powdered gloves are limited to 15 mg/dm².

Who manufactures sterile surgical gloves in India for export?

Vibama Surgicals, in Surat, makes sterile latex surgical gloves (REF PPE-003), sterile nitrile latex-free surgical gloves (REF PPE-016) and gynaecological long gloves (REF PPE-004) under a CDSCO manufacturing licence and an ISO 13485 certified quality system, with CE-marked products and a CoA per lot. See the gloves range or contact the export team.